Recovery of drugs of abuse from the Immunalysis Quantisal oral fluid collection device.
J Anal Toxicol. 2006 Oct;30(8):614-6
Authors: Quintela O, Crouch DJ, Andrenyak DM
Abstract
Drug recovery from a new oral fluid collection device was assessed. The evaluation was performed in vitro at three physiologically relevant concentrations for the following substances: amphetamine, methamphetamine, morphine, codeine, cocaine, benzoylecgonine, methadone, oxazepam, and Delta(9)-tetrahydrocannabinol (THC). Drug-free and drug-fortified controls were prepared and their concentration verified by liquid chromatography-tandem mass spectrometry. Aliquots of the controls were then "collected" with the device (n=3) using the manufacturer's recommended procedure. Collected samples were stored for 12 h to simulate shipping before analysis. Fresh, non-"collected" aliquots of each pool (n=3) were concurrently analyzed. The drug recoveries from the Quantisal were expressed as a mean percentage of the concurrently analyzed aliquots that were not subjected to device collection. Recoveries for oxazepam exceeded 97%, for amphetamine and methamphetamine exceeded 93%, and for opioids all exceeded 91%. The recoveries of cocaine were >91% and >82% for its polar metabolite, benzoylecgonine. Especially noteworthy was the recovery of THC from the Quantisal collector (81.3-91.4%). When compared with recoveries reported from other collection devices, the Quantisal was clearly superior.
PMID: 17132261 [PubMed - indexed for MEDLINE]